Shenzhen Shijia Instruments Co., Ltd.

Products

Specializing in spectrum analyzers, network analyzers, signal generators, LCR meters, oscilloscopes, power supplies, probes, digital multimeters, data acquisition systems, source measure units, power meters, battery testers, resistance testers and other high-frequency test instruments

E4982A LCR Meter

  • ModelE4982A
  • AttributesLCR Meter and Impedance Measurement Products
  • Category KEYSIGHT > LCR Meters & Impedance Measurement Products
  • BrandKEYSIGHT
  • DescriptionThe Keysight E4982A precision impedance analyzer covers a wide frequency range of 1 MHz to 3 GHz for high-precision RF impedance measurement, suitable for component, material and semiconductor characterization analysis.

Detailed Introduction Technical Documents

Introduction

Keysight E4982A Precision Impedance Analyzer, with a frequency range from 1 MHz to 3 GHz, provides high-precision RF impedance measurement and is suitable for component, material and semiconductor testing.

Features

 

Wide frequency range: 1 MHz to 3 GHz (optional 300 MHz/500 MHz/1 GHz/3 GHz)

Basic impedance accuracy: ±0.51% to ±4.32% (depending on frequency and level) 

Measurement parameters: |Z|, |Y|, θ, L, C, R, X, G, B, D, Q, Rdc 

Oscillator level: -33 dBm to +1 dBm 

Three measurement speed modes (Mode 1/2/3)

List sweep: up to 8 lists, 201 points per list

Calibration compensation:OPEN/SHORT/LOAD, low loss capacitor, port extension

Built-in DC resistance (Rdc) measurement

10.4 inch color LCD with XGA resolution

 

Applications

 

RF passive component (inductor, capacitor, filter) testing

Impedance analysis for semiconductor devices (varactor diodes, PIN diodes)

Characteristic measurement of dielectric materials and magnetic materials

Impedance matching for antennas and RFID coils

Automated sorting on production lines

 

Technical Specifications

Parameter

Specification

Frequency Range

1 MHz – 3 GHz (optional up to 300 MHz / 500 MHz / 1 GHz)

Frequency Resolution

100 kHz (Firmware B.02.20 and above)

Measurement Parameters

|Z|, |Y|, θ, L, C, R, X, G, B, D, Q, Rdc

Impedance Measurement Range

0.1 Ω – 10 kΩ (typical)

Basic Impedance Accuracy

0.51% (1–100 MHz, 1 dBm) ~ 4.32% (1.8–3 GHz, 1 dBm)

Oscillator Level

-33 dBm to +1 dBm

Level Accuracy

±0.5 dB (Typical)

Measurement Speed Mode

Mode 1 (Short), Mode 2 (Medium), Mode 3 (Long/High Precision)

Averaging Factor

1 – 100

List Sweep

201 points per table, up to 8 tables

Rdc Measurement

0.1 Ω – 100 Ω, 1 mΩ resolution

Calibration Type

OPEN/SHORT/LOAD, Low-loss Capacitor, Port Extension


Accessories

Standard: Test head (3.5 mm female), power cord, user manual, calibration certificate, N-type port protection cap

Optional: Frequency option (-032/052/102/302), 3.5 mm to 7 mm adapter, Handler interface, rack mount kit


Interfaces

Standard configuration: GPIB (IEEE-488), USB (USBTMC-USB488), LAN (10/100/1000 BaseT), USB Host, VGA video output, external trigger input, 10 MHz reference input/output (BNC)

Option: Handler interface (36-pin Centronics)



Physical Specifications

 

Dimensions (W×H×D): approx. 432 mm × 426 mm × 145 mm (main unit)

Weight: 13 kg for main unit, 250 g for test head